3 in 1 Ultrasound System
A/B Scan and Pachymetry
The new Compact Touch benefits from a new generation of 15 MHz B probe with a resolution increased by 30%. If allows for a better visualization of the eye structures and the orbit hence a better diagnosis.
Small in size, this probe benefits from an excellent ergonomics to facilitate its handling and use.
A weight now less than 4kg, Compact Touch has never been so well named, since its dimensions have been significantly reduced to fit better in every working environment
To further increase working ergonomics around the patient, its VESA fixation system offers the possibility to be mounted either:
The IOL calculation function allows comparison between different IOL types and calculation formulae. A total of 12 calculation formulae is available including post-refractive surgery formulae. The IOL implant calculation is done at 0.25D.
Essential for glaucoma diagnosis and for refractive surgery, Compact Touch offers several measuring modes and corneal maps with an accuracy of ± 5 microns.
Intraocular pressure can be adjusted thanks to built-in I0P correction calculation tables.
Published on 18 octobre 2017 / Quantel Medical Introduces Next-Generation Compact Touch Ophthalmic Ultrasound Platform at ESCRS2017
In this video, Robin Agopian (Ultrasound Product Manager in Quantel Medical) summarizes the improvements of the new generation of Compact Touch: a 15 MHz B prob
Quantel Medical Introduces Next-Generation Compact Touch Ophthalmic Ultrasound Platform at ESCRS2017
Mise en ligne le 18 octobre 2017 / Quantel Medical Introduces Next-Generation Compact Touch Ophthalmic Ultrasound Platform at ESCRS2017
In this video, Robin Agopian (Ultrasound Product Manager in Quantel Medical) summarizes the improvements of the new generation of Compact Touch: a 15 MHz B prob
New Compact Touch - Compact in Design, Definitive in Diagnosis
Mise en ligne le 05 octobre 2017 / New Compact Touch - Compact in Design, Definitive in Diagnosis
Compact Touch - Diseño Compacto, Diagnósticos Precisos
Mise en ligne le 06 octobre 2017 / Compact Touch - Diseño Compacto, Diagnósticos Precisos
We present a retrospective study of 1000 consecutive patient referred to Eplore Vision Center for IOL calculation. B Mode biometry is performed patient reclined, open eyes without anesthetic drop...
By Michel Puech, MD
Central corneal thickness influences intraocular pressure (IOP) measurement. We examined the central corneal thickness of subjects in the Ocular Hypertension Treatment Study (OHTS) and determined if central corneal thickness is related to race...
By James D. Brandt, MD; Juila A. Beiser, MS; Michael A. Kass; MD; Mac O. Gordon, PhD; and the Ocular Hypertension Treatment Study (OHTS) Group
Preoperative calculation of intraocular lens (IOL) refractive power is based on a number of mathematical formulas. These methods are usually derived from regression statistics applied to the postoperative refractive outcome of cataract surgery. In addition, some calculation techniques were developed from theoritical models of the eye’s optics…
By Olivier Bergès, MD; Michel Puech, MD; Michael Assouline, MD, PhD; Luc Letenneur, PhD; and Marc Gastellu-Etchegorry, MD
General diagnosis probe
15 MHz Probe for eye and orbit examination. Automatic biometry readings obtained from a B mode image allowing axial length measurements for difficult to mesure long eyes or posterior staphyloma.
Transducer frequency: | 15 MHz | Focus: | 24 mm (0.94”) |
|
Angle of exploration: | 50° | Axial resolution: | 115 μm | |
Depth of exploration: | 60 mm (2.36") |
Lateral resolution: Frame rate acquisition: |
400 μm up to 16 Hz |
|
Axial length measurements probe
11 MHz Probe | |
Adjustable gain: | 20 to 110 dB |
Time Gain Control (TGC): | 0 to 30 dB |
Transducer frequency: | 11 MHz |
Tip diameter: | 6 mm (0.23”) |
Electronic resolution: | 0.04 mm (0.002”) |
Depth: | 60 mm (2.4'') on 1536 points |
Contact and immersion techniques compatible. | |
Aiming beam: | LED |
Axial length measurements | |||
Ultrasound propagation velocity adjustable per segment (anterior chamber, lens, vitreous) and IOL and vitreous material. | |||
Built-in pattern recognition: | dense, phakic, aphakic, PMMA, acrylic and silicone material for pseudo-phakic eye types | ||
Automatic calculation of standard deviation and average total length (series of 10 measurements) | |||
Acquisition modes: | automatic, auto + save, manual | ||
Automatic detection of scleral spike |
Axial length measurements probe
Corneal thickness measurement
Specifications are subject to change without notice.
©2019. Quantel Medical, Compact Touch™ is a registered trademark of Quantel Medical. All rights reserved.
Ce site contient des informations réservées aux professionnels de santé.
Les contenus de ce site n'ont pas été validés pour un usage grand public. Afin d'accéder à ces pages vous devez confirmer que vous êtes un professionnel de santé ou assimilé.
Êtes-vous un professionnel de santé ou assimilé ?
Vous allez être redirigé vers un site internet dont la langue est différente du site actuel.
Cliquez sur "Rester sur l'actuel site internet" ou "Continuer vers le nouveau site internet"
Vous allez être redirigé vers un site internet différent de celui sur lequel vous naviguez jusqu'à présent.
Cliquez sur "Rester sur l'actuel site internet" ou "Continuer vers le nouveau site internet"
Ce site contient des informations réservées aux professionnels de santé.
Les contenus de ce site n'ont pas été validés pour un usage grand public. Afin d'accéder à ces pages vous devez confirmer que vous êtes un professionnel de santé ou assimilé.
Êtes-vous un professionnel de santé ou assimilé ?
Ce site contient des informations réservées aux professionnels de santé.
Les contenus de ce site n'ont pas été validés pour un usage grand public. Afin d'accéder à ces pages vous devez confirmer que vous êtes un professionnel de santé ou assimilé.
Êtes-vous un professionnel de santé ou assimilé ?