Global therapeutic solutions in ophthalmology
Patient Education
Aviso
A modular Ultrasound platform
UBM
High-frequency
A/B-scan
Technical information - A scan specifications Option
- Frequency: 11 MHz
- Diameter of tip: 6 mm (0.23")
- Method: contact, immersion
- Focus beam: LED or laser pointer*
- Electronic resolution: 0.04 mm (0.002)
- Depth: 45 mm (1.77")
- Points on X: 1536
- Level on Y: 256
- Measurement accuracy: ± 0.04 mm (0.002")
- Internal memory for saving data on the last 10 patients
- 5 different user profiles
- The following measurements are taken at each examination: anterior chamber, lens, vitreous body and total length
- Automatic detection of the scleral spike
- Propagation velocity; specific and adjustable for each tissue media
- 6 scan modes: normal, dense cataract/long eye, aphakic, pseudophakic in PMMA, acrylic and silicone
- 2 measurement techniques: contact or immersion
- 6 implant calculation formulae: SRK-T, SRK II, HOLLADAY, BINKHORST-II, HOFFER-Q, HAIGIS.
- 6 post-refractive formulae: history derived method, Refraction derived, contact lens method, ROSA regression, SHAMMAS regression and double K/SRK-T